The pigtails are specially designed to meet the requirements of NFPA for Hospital medgas manifold systems. The major feature being NFPA’s requirement for permanent affixed ends to avoid cross connection through modification or switching of the nut and nipple to hook up a different gas than a system was designed for. The other requirement is a mandatory copper pigtail for use in high pressure oxygen service. Some states also require that because Nitrous Oxide is an oxidizer, copper is also required.
- Black Urethane perforated cover
- 3/8" ID Nylon Core
- 4,000 PSI Working Pressure
- Permanent non removable CGA fitting ends per NFPA code
Unit of Measure
1
Item # |
Item Name |
CGA Standard |
Application |
Use with |
Inlet Thread Size |
Outlet Thread Size |
---|---|---|---|---|---|---|
PFSL-320-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 320 | N/A Carbon Dioxide | N/A Dewar Manifold System | N/A Female Nut | N/A Female Nut |
PFSL-326-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 326 | N/A Nitrous Oxide | N/A Dewar Manifold System | N/A Female Nut | N/A Female Nut |
PFSL-540-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 540 | N/A Oxygen | N/A Dewar Manifold System | N/A Female Nut | N/A Female Nut |
PFSL-580-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 580 | N/A Helium Nitrogen | N/A Dewar Manifold System | N/A Male Nut | N/A Male Nut |
Unit of Measure
1