The pigtails are specially designed to meet the requirements of NFPA for Hospital medgas manifold systems. The major feature being NFPA’s requirement for permanent affixed ends to avoid cross connection through modification or switching of the nut and nipple to hook up a different gas than a system was designed for. The other requirement is a mandatory copper pigtail for use in high pressure oxygen service. Some states also require that because Nitrous Oxide is an oxidizer, copper is also required.
- Black Urethane perforated cover
- 3/8" ID Nylon Core
- 2,250 PSI Working Pressure
- Permanent non removable CGA fitting ends per NFPA code
Unit of Measure
1
Item # |
Item Name |
CGA Standard |
Application |
Inlet Thread Size |
Outlet Thread Size |
Use with |
---|---|---|---|---|---|---|
PFSL-320-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 320 | N/A Carbon Dioxide | N/A Female Nut | N/A Female Nut | N/A Dewar Manifold System |
PFSL-326-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 326 | N/A Nitrous Oxide | N/A Female Nut | N/A Female Nut | N/A Dewar Manifold System |
PFSL-540-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 540 | N/A Oxygen | N/A Female Nut | N/A Female Nut | N/A Dewar Manifold System |
PFSL-580-72 | N/A 72" Liquid Cylinder Gas Use Pigtails | N/A 580 | N/A Helium Nitrogen | N/A Male Nut | N/A Male Nut | N/A Dewar Manifold System |
Unit of Measure
1